KAN SAYIM CİHAZLARI

Parameters35 reportable parameters: WBC, Neu#, Lym#, Mon#, Eos#, Bas#, IG#, Neu%, Lym%, Mon%, Eos%, Bas%, IG%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW, PCT, P-LCR, P-LCC, IPF, RET#, RET%, LFR, MFR, HFR, IRF, RHE
29 research parameters
Two 3D scattergrams & Eight 2D scattergrams & Three histograms
PrinciplesCBC+DIFF/RET: Semi-conductor Laser scattering & Fluorescent method & Flow cytometry method
WBC/RBC/PLT counting: Impedance method
HGB: Cyanide-free colorimetry method
ReagentPlasma Diluent-N (20 L); Plasma Diluent-R (1000 mL); Lyse-G (500 mL); Lyse-D (1000 mL); EH 6100 FDN-D (22 mL); EH 6100 FDN-R (12 mL)
Sample ModeVenous whole blood, capillary blood
RepeatabilityWBC ≤2.5%(4.0-15.0 x 109/L)
RBC ≤1.5%(3.5-6.0 x 1012/L)
HGB ≤1.0%(110-180g/L)
HCT ≤1.5%(30.0-50.0)
MCV ≤1%(70-120fL)
PLT ≤4.0%(100-500x 109/L)
Display12.1 inches color screen
Dimension665mm(W)-870mm(H)-820mm(D)
ThroughputUp to 100 tests per hour
Sample VolumeCBC: 20ul; CBC+DIFF: 30ul; CBC+DIFF+RET: 35ul (Whole blood)
StorageUp to 150,000 records
Linearity RangeWBC: 0.00-500 x 109/L
RBC: 0.00-8.60 x 1012/L
HGB: 0-260g/L
HCT: 0.0-75.0 %
PLT: 0-5000 x 109/L
HCT: 0-30.0 %
Net Weight100 kg
Data TransmissionBi-directional LIS supported
Power RequirementAC 100V ~ 240V (50-60Hz)
Parameters25 reportable parameters: WBC, Neu#, Lym#, Mon#, Eos#, Bas#, Neu%, Lym%, Mon%, Eos%, Bas%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW, PCT, P-LCR, P-LCC
4 research parameters: ALY#, ALY%, LIC#, LIC%
3 histograms for WBC, RBC and PLT
One 3D scattergrams and three 2D scattergrams for WBC differention
PrinciplesElectrical Impedance method for WBC/BAS count and RBC/PLT tests
Cyanide free colorimetry for HGB test
Laser-based flow cytometry for WBC differentiation
ReagentPlasma Diluent 590 (20 L); Lyse EH 1 (500 mL); Lyse EH 2 (500 mL); Lyse EH 3 (1000 mL)
Sample ModeWhole blood, Capillary whole blood and Pre-diluted
RepeatabilityWBC ≤2.0%(4.0-15.0 x 109/L)
RBC ≤1.5%(3.5-6.0 x 1012/L)
HGB ≤1.5%(110-180g/L)
MCV ≤1%(70-120fL)
PLT ≤6.0%(100-149x 109/L)
≤4.0%(150-500 x 109/L)
Dimension650mm(W)-550mm(H)-610mm(D)
ThroughputUp to 90 tests per hour
Sample Volume20 µL
StorageUp to 100,000 records
Linearity RangeWBC: 0.00-300 x 109/L
RBC: 0.00-8.50 x 1012/L
HGB: 0-250g/L
PLT: 0-3000 x 109/L
HCT: 0 -67%
Net Weight57.5 kg
Data TransmissionSupport bi-directional LIS
Power Requirement100V ~ 240V, 50/60Hz, ≤ 250VA
Parameters21 reportable parameters: WBC, Gran#, Lym#, Mid#, Gran%, Lym%, Mid%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW, PCT, P-LCR, P-LCC
3 histograms for WBC, RBC and PLT
PrinciplesElectrical Impedance method for WBC, RBC and /PLT tests
Cyanide free colorimetry for HGB test
ReagentPlasma Diluent EH ( 20 L); Stromalyse EH (500 mL)
Sample ModeWhole blood ,Capillary blood and Pre-diluted
RepeatabilityWBC ≤2.0%(7.0-15.0 x 109/L)
≤3.5%(4.0-6.9 x 109/L)
RBC ≤1.5%(3.50-6.50 x 1012/L)
HGB ≤1.5%(100-180g/L)
MCV ≤0.5%(70.0-110fL)
PLT ≤4.0%(150-500x 109/L)
≤5.0%(100-149 x 109/L)
Display10.4 inches TFT touch screen
Dimension364mm(W)-477mm(H)-417mm(D)
ThroughputUp to 60 tests per hour
Sample Volume9 µL
StorageUp to 50,000 records
Linearity RangeWBC: 0.00-300 x 109/L
RBC: 0.00-8.50 x 1012/L
HGB: 0-250g/L
PLT: 0-3000 x 109/L
HCT: 0.0 -67%
Net Weight26 kg
Data TransmissionBi directional LIS with LAN port
Power Requirement100V-240V ~, 50/60Hz